1 January 1981 Low And Moderate Resolution Spectroscopy With An Rca Charge-Coupled Device (CCD)
J. B. Oke
Author Affiliations +
Abstract
An RCA buried channel thinned 320x 512 CCD has recently been put into operation as the detector on the blue side of a new double spectrograph used at the Cassegrain focus of the Hale 5.08-meter telescope. The chip is used in a vacuum dewer at a temperature of -132°C. The electronics are essentially the same as those used by Westphal and Gunn for the Texas Instruments 800x 800 CCD's. The readout noise over most of the chip is 46 electrons per pixel, but on one side of the device there are columns with readout noise three times as high as this. The dark current is approximately 85 electrons per pixel per hour and for most purposes can be neglected. When used in a spectroscopic mode with an f/1.5 beam the CCD shows interference patterns with an amplitude of 5 to 20 percent. This pattern occurs most clearly above 5000 Å. It appears to be stable and can be removed by using standard flat-field exposures. At signal levels of a few hundred electrons there are some charge transfer problems. The quantum efficiency between 4000 and 6500 Å is very high and the efficiency at 3260 Å is about 20 percent of that at 4200 Å. Above 6500 A the sensitivity begins to drop steadily.
© (1981) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. B. Oke "Low And Moderate Resolution Spectroscopy With An Rca Charge-Coupled Device (CCD)", Proc. SPIE 0290, Solid-State Imagers for Astronomy, (1 January 1981); https://doi.org/10.1117/12.965835
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Charge-coupled devices

Quantum efficiency

Stars

Electrons

CCD cameras

Astronomy

Electronics

Back to Top