Paper
4 November 1981 Noise And Material Residues In Dual-Energy Basis Decomposition Radiography
Leonard A. Lehmann, Albert Macovski
Author Affiliations +
Abstract
Intervening tissues in projection radiography often seriously interfere with the observability of other low contrast structures. Dual energy techniques resolve this difficulty by removing a selected substance from the image, enhancing the relative contrast of other materials. Undesired materials will remain; these residues, as well as quantum noise, limit the accuracy of material isolation. Parameterizations of the attenuation coefficient lead to simple expressions for the signal-to-noise ratio (SNR) in the dual energy radiograph. The SNR depends on the absorption path, the imaging energies, and the cancelled material. The magnitude of material residue depends only on the atomic numbers and may be an important consideration in the selection and development of new contrast agents.
© (1981) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Leonard A. Lehmann and Albert Macovski "Noise And Material Residues In Dual-Energy Basis Decomposition Radiography", Proc. SPIE 0314, Digital Radiography, (4 November 1981); https://doi.org/10.1117/12.933036
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Bone

Radiography

Signal attenuation

Iodine

Signal to noise ratio

Dual energy imaging

Aluminum

RELATED CONTENT

Iodine Imaging Using Three Energy Spectra
Proceedings of SPIE (November 04 1981)
Limitations To Accuracy In Dual Energy Digital Fluoroscopy
Proceedings of SPIE (November 04 1981)
The Energy Factor In Radiographic Imaging
Proceedings of SPIE (November 01 1982)

Back to Top