Paper
29 December 1982 Calibration Of A Digital Radiography System For Quantitative Studies
C. G. Shaw, D. A. Bassano, Z. D. Grossman
Author Affiliations +
Proceedings Volume 0347, Application of Optical Instrumentation in Medicine X; (1982) https://doi.org/10.1117/12.933818
Event: Application of Optical Instrumentation in Medicine X, 1982, New Orleans, United States
Abstract
The generation of contrast signals through digital mask-mode subtraction is studied with an emphasis on the Low contrast situations. The effects of beam hardening, radiation scattering, and veiling glare are discussed. The linearity between the contrast signal and the iodine concentration is tested at contrast levels from 3.7 mg/cm3 to 370 mg/cm3 of iodine in a plastic tube of 2.7 urn in diameter. A xenon perfusion study of a canine brain is presented.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
C. G. Shaw, D. A. Bassano, and Z. D. Grossman "Calibration Of A Digital Radiography System For Quantitative Studies", Proc. SPIE 0347, Application of Optical Instrumentation in Medicine X, (29 December 1982); https://doi.org/10.1117/12.933818