1 January 1987 Fast Magnetic Resonance Imaging Using Spiral Trajectories
Mark J. Blum, Michael Braun, Dov Rosenfeld
Author Affiliations +
Abstract
Acquisition times in magnetic resonance imaging (MRI) are typically in the order of minutes. For an image of 256 x 256 pixels, the standard Fourier reconstruction technique used in most commercial imaging systems requires 256 separate free induction decay (FID) signals. While the FID signal itself is of relatively short duration, the successive FID signals are separated by long delays, of the order of seconds, to permit substantial relaxation of the signal before the next excitation. The resultant long acquisition times give rise to motion artefacts, preclude dynamic imaging and keep the patient throughput low. In this paper, we investigate a fast imaging scheme which uses spiral trajectories in the spatial frequency domain. The entire domain can be sampled in a short time, requiring as few as one FID acquisition. The scheme requires time varying gradients having the form of a ramped sinusoid. Several reconstruction methods are considered for forming the image from the spatial frequency domain data. The possibility of using multiple spirals to deal with the rapid decay of the FID signal is also examined in detail.
© (1987) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mark J. Blum, Michael Braun, and Dov Rosenfeld "Fast Magnetic Resonance Imaging Using Spiral Trajectories", Proc. SPIE 0767, Medical Imaging, (1 January 1987); https://doi.org/10.1117/12.966978
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Magnetic resonance imaging

Spatial frequencies

Fourier transforms

Medical imaging

Raster graphics

Computer simulations

Image quality

Back to Top