27 June 1988 Blood Flow Determination From Fluoroscopic Image Sequences
M. Hartmann, M. A. Simons, B. Yih, R. A. Kruger
Author Affiliations +
Abstract
In this paper a method to quantitate blood flow from contrast injections into a vessel was evaluated using a straight tube, constant flow phantom. Grey scale values integrated over multiple regions-of-interest (ROI), placed contiguously along a "vessel" segment, were used to generate multiple time-density curves. Temporal shifts in the time-density curves for each ROI were determined using a cross-correlation technique. Regional volume estimates were made using a curve fitting approach. Linear regression of the regional volume and temporal shift estimates was used to estimate flow within the vessel segment. Velocity estimates were made in a similar way. Currently, velocity estimations are significantly better than flow estimations. Errors in the diameter estimates remain as the main cause of flow inaccuracies.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. Hartmann, M. A. Simons, B. Yih, and R. A. Kruger "Blood Flow Determination From Fluoroscopic Image Sequences", Proc. SPIE 0914, Medical Imaging II, (27 June 1988); https://doi.org/10.1117/12.968658