27 January 2017 Monitoring of nitrous acid (HONO) by external-cavity quantum cascade laser-based off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS)
Hongming Yi, Rabih Maamary, Xiaoming Gao, Markus W. Sigrist, Eric Fertein, Weidong Chen
Author Affiliations +
Abstract
In this presentation, we report on the high-sensitivity and high-selectivity measurement of HONO by off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) in a very small gas sample volume (of ~ 40 mm3) resulting in a ultrashort residence time of less than 10 ms (compared to ~ 7 min for a conventional 210 m multipass cell or ~ 10-min. integration time for currently used chemical analytical instruments). A minimum detection limit of 66 ppbv (1σ) HONO was achieved at 70 mbar using a laser output power of 50 mW and 1 s integration time. This MDL was down to 7 ppbv at the optimal integration time of 150 s.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hongming Yi, Rabih Maamary, Xiaoming Gao, Markus W. Sigrist, Eric Fertein, and Weidong Chen "Monitoring of nitrous acid (HONO) by external-cavity quantum cascade laser-based off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS)", Proc. SPIE 10111, Quantum Sensing and Nano Electronics and Photonics XIV, 1011137 (27 January 2017); https://doi.org/10.1117/12.2251096
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Quantum cascade lasers

Photoacoustic spectroscopy

Modulation

Absorption

Sensors

Signal detection

Chemical analysis