You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
3 March 2017Patient-specific atrium models for training and pre-procedure surgical planning
Minimally invasive cardiac procedures requiring a trans-septal puncture such as atrial ablation and MitraClip® mitral
valve repair are becoming increasingly common. These procedures are performed on the beating heart, and require
clinicians to rely on image-guided techniques. For cases of complex or diseased anatomy, in which fluoroscopic and
echocardiography images can be difficult to interpret, clinicians may benefit from patient-specific atrial models that can
be used for training, surgical planning, and the validation of new devices and guidance techniques.
Computed tomography (CT) images of a patient’s heart were segmented and used to generate geometric models to create
a patient-specific atrial phantom. Using rapid prototyping, the geometric models were converted into physical
representations and used to build a mold. The atria were then molded using tissue-mimicking materials and imaged using
CT. The resulting images were segmented and used to generate a point cloud data set that could be registered to the original
patient data. The absolute distance of the two point clouds was compared and evaluated to determine the model’s accuracy.
The result when comparing the molded model point cloud to the original data set, resulted in a maximum Euclidean
distance error of 4.5 mm, an average error of 0.5 mm and a standard deviation of 0.6 mm.
Using our workflow for creating atrial models, potential complications, particularly for complex repairs, may be accounted
for in pre-operative planning. The information gained by clinicians involved in planning and performing the procedure
should lead to shorter procedural times and better outcomes for patients.
The alert did not successfully save. Please try again later.
Justin Laing, John Moore, Daniel Bainbridge, Maria Drangova, Terry Peters, "Patient-specific atrium models for training and pre-procedure surgical planning," Proc. SPIE 10135, Medical Imaging 2017: Image-Guided Procedures, Robotic Interventions, and Modeling, 101351A (3 March 2017); https://doi.org/10.1117/12.2249693