28 July 2017 Study of extracerebral contamination for three cerebral oximeters by Monte Carlo simulation using CT data
Author Affiliations +
Abstract
Using multidetector computed tomography, thicknesses of bone squame and soft tissues of human head were assessed. MC simulation revealed impropriety of source-detector separation distances for 3 oximeters, which can cause extracerebral contamination.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
A. P. Tarasov, A. I. Egorov, and D. A. Rogatkin "Study of extracerebral contamination for three cerebral oximeters by Monte Carlo simulation using CT data", Proc. SPIE 10411, Clinical and Preclinical Optical Diagnostics, 104110R (28 July 2017); https://doi.org/10.1117/12.2281276