Presentation + Paper
7 February 2018 Comparison of a novel high-power blue diode laser (λ=442 nm) with Ho:YAG (λ=2100 nm), Tm fiber (λ=1940 nm), and KTP (λ=532 nm) lasers for soft tissue ablation
Victoriya Vinnichenko, Anastasiya Kovalenko, Valeriya Arkhipova, Ilya Yaroslavsky, Gregory Altshuler, Valentin Gapontsev
Author Affiliations +
Abstract
Three lasers were directly compared, including the Ho:YAG laser (λ = 2100 nm), Tm fiber laser (λ = 1940 nm) operating in 3 different modes (CW, regular pulse, and super pulse), and blue diode laser (λ = 442 nm) for vaporization and coagulation efficiency for treating blood-rich soft tissues, ex vivo, in a porcine kidney model at quasi-contact cutting in water. In addition, experimental results were compared with published data on performance of KTP laser (λ = 532 nm) at similar experimental settings (Power = 60 W and cutting speed = 2 mm/s). Tm fiber laser in pulsed mode and blue laser produced highest vaporization rates of 3.7 and 3.4 mm3/s, respectively. Tm fiber laser (in both CW and pulsed modes) also produced the largest coagulation zone among the laser sources tested. A carbonization zone was observed for Tm fiber laser in CW and pulsed modes, as well as for the blue diode laser. Tm fiber laser in super-pulse mode and Ho:YAG laser both resulted in irregular coagulation zones without carbonization. Comparison with known data for KTP laser revealed that tissue effects of the blue laser are similar to that of the KTP laser. These results suggest that the combination of the two lasers (Tm fiber and blue diode) in one system may achieve high cutting efficiency and optimal coagulation for hemostasis during surgical treatment. Ex vivo testing of the combined system revealed feasibility of this approach. The combination of the CW Tm fiber laser (120W) and the blue diode laser (60W) emitting through a combination tip were compared with CW 120 W Tm fiber laser alone and 120 W Ho:YAG laser. Vaporization rates measured 34, 28, and 6 mm3/s, and coagulation zones measured 0.6, 1.3, and 1.7 mm, respectively. A carbonization zone was only observed with CW Tm fiber laser. The vaporization rate of combined CW Tm fiber laser / blue diode laser was comparable to published data for KTP laser for equivalent total power. Thus, high-power blue diode laser, Tm fiber laser, and their combination may provide an alternative to conventional Ho:YAG and KTP lasers for applications in urology and other surgical fields.
Conference Presentation
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Victoriya Vinnichenko, Anastasiya Kovalenko, Valeriya Arkhipova, Ilya Yaroslavsky, Gregory Altshuler, and Valentin Gapontsev "Comparison of a novel high-power blue diode laser (λ=442 nm) with Ho:YAG (λ=2100 nm), Tm fiber (λ=1940 nm), and KTP (λ=532 nm) lasers for soft tissue ablation", Proc. SPIE 10468, Therapeutics and Diagnostics in Urology 2018, 104680F (7 February 2018); https://doi.org/10.1117/12.2291103
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fiber lasers

Thulium

Semiconductor lasers

Continuous wave operation

Laser cutting

Ferroelectric materials

Laser tissue interaction

RELATED CONTENT

OCT/OCTA-guided laser ablation for tumor treatment
Proceedings of SPIE (October 27 2021)
In vitro testing of dual-mode thulium microsurgical laser
Proceedings of SPIE (February 09 2012)
Holmium-laser endopyelotomy
Proceedings of SPIE (May 02 1994)
Use of the 2 µm cw laser as addition and...
Proceedings of SPIE (March 23 2007)

Back to Top