You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
9 March 2018Assessment of image quality parameters of a novel micro-CT system compared to a conventional CT geometry
A novel fourth generation micro-CT (WATCH-CT) with a unique scanning geometry, that collects parallel projections from a standard x-ray source without the requirement to interpolate or rebin the data, is studied and evaluated for its imaging qualities and performance characteristics. For a comparative analysis of the WATCH micro-CT system and the conventional CT geometry, the local noise power spectrum and the modulation transfer function is derived from the same initial parameters. The spatial resolution (MTF), characterized by the response of the system, is determined by the MTF derived by the oversampling method. The calculations involve varying the parameters like the region of evaluation (ROE) position, FOV magnification, angular sampling, pixel size, filtration and reconstruction algorithm to provide an extensive analogy between these systems. The spatial resolution of the scanning geometries is evaluated and compared. The MTF curves illustrate a higher relative resolving capacity for the WATCH micro-CT compared to the conventional geometries which is due to the characteristics of this unique geometry. The WATCH system exhibits higher resolutions explicitly at the regions away from the center. The NPS curves of WATCH geometry shows higher noise content in comparison to the conventional geometry.
The alert did not successfully save. Please try again later.
K. Kumar, N. Saeid Nezhad, B H. Mueller, O. Tischenko , C. Hoeschen, "Assessment of image quality parameters of a novel micro-CT system compared to a conventional CT geometry," Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105732L (9 March 2018); https://doi.org/10.1117/12.2293774