Paper
12 January 2018 Research on the measurement of the ultraviolet irradiance in the xenon lamp aging test chamber
Muyao Ji, Tiecheng Li, Fangsheng Lin, Dejin Yin, Weihai Cheng, Biyong Huang, Lei Lai, Ming Xia
Author Affiliations +
Abstract
This paper briefly introduces the methods of calibrating the irradiance in the Xenon lamp aging test chamber. And the irradiance under ultraviolet region is mainly researched. Three different detectors whose response wave range are respectively UVA (320~400nm), UVB (275~330nm) and UVA+B (280~400nm) are used in the experiment. Through comparing the measuring results with different detectors under the same xenon lamp source, we discuss the difference between UVA, UVB and UVA+B on the basis of the spectrum of the xenon lamp and the response curve of the detectors. We also point out the possible error source, when use these detectors to calibrate the chamber.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Muyao Ji, Tiecheng Li, Fangsheng Lin, Dejin Yin, Weihai Cheng, Biyong Huang, Lei Lai, and Ming Xia "Research on the measurement of the ultraviolet irradiance in the xenon lamp aging test chamber", Proc. SPIE 10621, 2017 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems, 106211L (12 January 2018); https://doi.org/10.1117/12.2295285