Presentation + Paper
16 July 2018 Commissioning of cryogenic preamplifiers for SAPHIRA detectors
Author Affiliations +
Abstract
SAPHIRA detectors, which are HgCdTe linear avalanche photodiode arrays manufactured by Leonardo, enable high frame rate, high sensitivity, low noise, and low dark current imaging at near-infrared wavelengths. During all University of Hawaii Institute for Astronomy lab testing and observatory deployments of SAPHIRA detectors, there was approximately one meter of cables between the arrays and the readout controllers. The output drivers of the detectors struggled to stably send signals over this length to the readout controllers. As a result, voltage oscillations caused excess noise that prevented us from clocking much faster than 1 MHz. Additionally, during some deployments, such as at the SCExAO instrument at Subaru Telescope, radio-frequency interference from the telescope environment produced noise many times greater than what we experienced in the lab. In order to address these problems, collaborators at the Australia National University developed a cryogenic preamplifier system that holds the detector and buffers the signals from its outputs. During lab testing at 1 MHz clocking speeds, the preamplifiers reduced the read noise by 45% relative to data collected using the previous JK Henriksen detector mount. Additionally, the preamplifiers enabled us to increase the clocking frequency to 2 MHz, effectively doubling the frame rate to 760 Hz for a full (320x256 pixel) frame or 3.3 kHz for a 128x128 pixel subarray. Finally, the preamplifiers reduced the noise observed in the SCExAO environment by 65% (to essentially the same value observed in the lab) and eliminated the 32-pixel raised bars characteristic of radio-frequency interference that we previous observed there.
Conference Presentation
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sean B. Goebel, Donald N. B. Hall, Shane Jacobson, Annino Vaccarella, Rob Sharp, Michael Ellis, and Izabella Pastrana "Commissioning of cryogenic preamplifiers for SAPHIRA detectors", Proc. SPIE 10709, High Energy, Optical, and Infrared Detectors for Astronomy VIII, 107090R (16 July 2018); https://doi.org/10.1117/12.2311513
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Cryogenics

Cameras

Adaptive optics

Telescopes

Signal detection

Electronics

RELATED CONTENT

A laser tomography test bed for extremely large telescopes
Proceedings of SPIE (August 04 2014)
Adaptive optics program at TMT
Proceedings of SPIE (July 10 2018)
Instrumentation for ESO's Extremely Large Telescope
Proceedings of SPIE (July 06 2018)
MAD: the ESO multiconjugate adaptive optics demonstrator
Proceedings of SPIE (February 07 2003)
Gemini instrument program
Proceedings of SPIE (August 16 2000)

Back to Top