You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
25 May 1989ROC Comparison of Analog Film to 512, 1024 and 2048 Line Digital Displays
A receiver operator characteristic (ROC) experiment designed to evaluate the diagnostic performance of 512, 1024 and 2048 monitors as compared to conventional analog films viewed on a lightbox was conducted. Fifty radiographs, 25 with and 25 without a simulated solitary nodule placed on the lung fields of an anthropomorphic chest phantom were used. These images were interpreted by six radiologists under the following six conditions: 512 and 1024 line resolution both with and without interactive contrast enhancement, and 2048 line resolution with contrast enhancement, and analog film on a conventional lightbox. As determined by the area under the ROC curve, which was the measure of diagnostic accuracy the 512 and 1024 line monitors performed below the analog film, the 2048 line monitor proved to be a viable alternative to the analog film.
The alert did not successfully save. Please try again later.
A. Hayrapetian, C. Morioka, H. K. Huang, R. Fiske, D. Aberle, M. I. Boechat, K. Brown, T. Hall, M. Milos, F. Tessler, "ROC Comparison of Analog Film to 512, 1024 and 2048 Line Digital Displays," Proc. SPIE 1092, Medical Imaging III: Image Processing, (25 May 1989); https://doi.org/10.1117/12.953295