Paper
1 March 2019 Demonstration of phase-assisted material decomposition with a Talbot-Lau interferometer using a single x-ray tube potential
E. R. Shanblatt, B. J. Nelson, S. Tao, S. Leng, C. H. McCollough
Author Affiliations +
Abstract
We present a proof-of-principle demonstration of material decomposition using a single X-ray tube potential (38 kVp + 0.2 mm Sn, for an effective energy around 27 keV) with a Talbot-Lau grating-based phase-contrast computed tomography (CT) system. We show good material separation of water and fat and an accurate quantitative measurement of isopropyl alcohol. This method utilizes the distinctiveness of both components of the refractive index, δ and β, and is promising for separating soft tissue materials that have similar attenuation values such as water and fat.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
E. R. Shanblatt, B. J. Nelson, S. Tao, S. Leng, and C. H. McCollough "Demonstration of phase-assisted material decomposition with a Talbot-Lau interferometer using a single x-ray tube potential", Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109482W (1 March 2019); https://doi.org/10.1117/12.2511806
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Refractive index

Water

Signal attenuation

X-rays

Tissues

Phase contrast

Phase measurement