1 March 2019 A practical analysis of scatter-to-primary ratio after beam hardening correction for x-ray CT
Hewei Gao, Shumei Jia, Geng Fu
Author Affiliations +
Abstract
Scatter and beam hardening degrade image quality for X-ray computed tomography (CT). Applying beam hardening correction before scatter being fully removed may be needed in some scenarios. In this paper, we theoretically analyze the outcome of beam hardening correction with scatter uncorrected. Using the effective spectrum of CT system, beam hardening correction function and its derivatives can be explicitly derived and used, along with a Taylor expansion, to approximate the change of scatter-to-primary ratio (SPR) after beam hardening correction. Two types of SPR compensations are compared using cone- and fan-beam scans of a CatPhan phantom on a tabletop cone-beam CT system, validating our analysis to be accurate in practical applications.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hewei Gao, Shumei Jia, and Geng Fu "A practical analysis of scatter-to-primary ratio after beam hardening correction for x-ray CT", Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 1094832 (1 March 2019); https://doi.org/10.1117/12.2513266