You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
1 March 2019A new test method to assess the representation of spiculated mass like targets in digital mammography and breast tomosynthesis
In this work we tested different materials for 3D printing of spiculated mass models for their incorporation into an existing 3D structured phantom for performance testing of FFDM and DBT. Counting the number of spicules as a function of dose was then evaluated as a possible extra test feature expressing conspicuity next to detectability. Seven printable materials were exposed together with a PMMA step wedge and material samples with known linear attenuation coefficient to determine PMMA equivalent thickness and linear attenuation coefficient, respectively. Next, two models of spiculated masses were created each with a different complexity in terms of number of spicules. The visibility of the number of spicules of a 3D printed spiculated mass model loosely placed in the phantom or embedded into two different printing materials was assessed for FFDM and DBT. Vero White pure was chosen as the most appropriate material for the printing of masses whereas Vero Clear and Tango+ were chosen as background materials. The visibility of spicules was best in the loose mass models and better in the background material Tango+ compared to Vero Clear. While the discrimination of the different spicules could be assessed in FFDM and DBT, as expected only a limited dose sensitivity was found for the visibility of spicules evaluated for the different background materials and at different beam energies.
The alert did not successfully save. Please try again later.
Elisabeth Salomon, Friedrich Semturs, Lesley Cockmartin, Michael Figl, Hilde Bosmans, Johann Hummel, "A new test method to assess the representation of spiculated mass like targets in digital mammography and breast tomosynthesis," Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109483Y (1 March 2019); https://doi.org/10.1117/12.2513464