6 September 2019 Retardance parameters measurements by employing a rotating polarizer-analyzer polarimeter
Author Affiliations +
Abstract
In this work, we proposed the usage of a rotating polarizer-analyzer polarimeter to retrieve retardance parameters of a sample by treating it as an elliptical retarder that can be characterized by three parameters: total retardance, fast axis orientation and an ellipticity related parameter. By employing the Mueller matrix approach, we developed the demodulation algorithm to retrieve each parameter with the association of the Fourier series analysis. We present experimental results considering dextrose concentrations diluted in distilled water scaling in a range from 30mg/dl to 45mg/dl. For comparison purposes, we employed the method of rotating an analyzer and compare the optical activity variation for both methods.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Humberto Macias-Mendoza, David I. Serrano-García, Jorge L. Flores, Guillermo García Torales, Azael Mora-Núñez, Héctor Santiago Hernández, Geliztle A. Parra Escamilla, and Francisco J. Cervantes Lozano "Retardance parameters measurements by employing a rotating polarizer-analyzer polarimeter", Proc. SPIE 11132, Polarization Science and Remote Sensing IX, 1113213 (6 September 2019); https://doi.org/10.1117/12.2529659