1 December 1989 Comparison Of Multivariate Calibration With PCR And PLS For Glucose Using Infrared Spectra Of Human Whole Blood
R. Marbach, H. M. Heise
Author Affiliations +
Proceedings Volume 1145, 7th Intl Conf on Fourier Transform Spectroscopy; (1989) https://doi.org/10.1117/12.969537
Event: Seventh International Conference on Fourier and Computerized Infrared Spectroscopy, 1989, Fairfax, VA, United States
Abstract
The calibration of glucose in whole EDTA blood was carried out using FT-IR spectra obtained with a micro-CIRCLE cell in the range of 1500 - 750 cm-1. The reference concentrations had been measured by the enzymatic glucose dehydrogenase method. The P-matrix approach was chosen applying the Partial-least-squares (PLS) and Principal component regression (PCR) algorithms. The optimum model for predictipns was obtained by PLS achieving an average prediction error of 19.8 mg/dL based on PRESS1 from crossvalidation. The calibration results for both algorithms are discussed. PLS is favoured on the grounds of numerical and statistical advantages.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
R. Marbach and H. M. Heise "Comparison Of Multivariate Calibration With PCR And PLS For Glucose Using Infrared Spectra Of Human Whole Blood", Proc. SPIE 1145, 7th Intl Conf on Fourier Transform Spectroscopy, (1 December 1989); https://doi.org/10.1117/12.969537
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Blood

Glucose

Error analysis

Statistical analysis

Spectroscopy

Magnesium

Back to Top