You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
22 May 2020Evaluation of breast compression using pressure distribution over a breast phantom and x-ray images in mammography
Breast compression in mammography is used to obtain suitable imaging. Compression reduces breast thickness and spreads the tissue, leading to improved image quality and reduced radiation dose. Breast compression is, however, uncomfortable for clients, and may often cause strong pain. In this study, we obtained pressure distribution and X-ray images of a
compressed breast simulated phantom using a pressure sensor. The purpose of this study was to visualize the pressure distribution and measure how much the phantom was spread by compression. The pressure sensor was placed on a breast support table, and the breast phantom was positioned in the medio-lateral oblique direction. Compression was applied in steps of 20 N from 40- 180 N. Indicated values of the pressure sensor and width of the compressed phantom were measured at each increased force step. Width differences from a non-compressed phantom were calculated at each applied compression force. The pressure distribution showed nonuniform pressure over the breast phantom, and high pressure in the juxta-thoracic region, which decreased toward the periphery. The width of the breast increased with increasing applied compression force in all positions, but the change became smaller after 140 N. By visualizing the pressure, we were able to evaluate the pressure distribution over all of the phantom.
The alert did not successfully save. Please try again later.
Hiroko Nishide, Tomomi Minemura, Fuyuka Morishita, "Evaluation of breast compression using pressure distribution over a breast phantom and x-ray images in mammography," Proc. SPIE 11513, 15th International Workshop on Breast Imaging (IWBI2020), 115132B (22 May 2020); https://doi.org/10.1117/12.2564350