15 February 2021 Radiographic imaging with a linear x-ray source and multi-row detector
Qinghao Chen, Shuang Zhou, Yuewen Tan, Tiezhi Zhang
Author Affiliations +
Conference Poster
Abstract
A tetrahedron beam CT (TBCT) benchtop system has been developed with linear array x-ray source with 48 focal spots in 4 mm spacing and a photon counting detector (PCD). The generated x-ray beams are collimated to fan beams by a multi-slot collimator and converge to the 6 mm wide multi-row photon counting detector. Due to its scatter rejection geometry, the scatter-to-primary ratio (SPR) is significantly reduced to 17% for fan beam comparing to 120% for cone beam in presence of a head phantom. We performed both 2D and 3D imaging study with the TBCT system. 2D radiography images have been obtained using shift-and-add method. Both analytical and iterative reconstruction methods were used in 3D image reconstruction. Head phantom and animal cadaver were scanned.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Qinghao Chen, Shuang Zhou, Yuewen Tan, and Tiezhi Zhang "Radiographic imaging with a linear x-ray source and multi-row detector", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 1159537 (15 February 2021); https://doi.org/10.1117/12.2582159