15 February 2021 Accuracy and precision of a structured light scanning system for acquiring 3D information of the breast under compression
F. Schampers, M. C. Pinto, K. Michielsen
Author Affiliations +
Conference Poster
Abstract
Realistic three-dimensional information about the compressed breast shape is required for the development of some image processing techniques and improvement of image quality in digital mammography and breast tomosynthesis (DBT). Our aim is to evaluate the precision and accuracy of the breast shape recorded by a structured light (SL) surface scanning system and to examine the similarity between these breast surface scans (BSS) and the depth information contained in the DBT images acquired concurrently, in a cranio-caudal view. Phantom tests were performed to examine the agreement between left- and right-side scans by measuring the estimated length and thickness. To evaluate accuracy, the Hausdorff distance was computed between the scanned phantom and the phantom’s known shape, and also between several measurements performed over time. The agreement between the BSS and the corresponding DBT data was assessed by comparing the width and chest-wall to nipple distance (CND) from both measurements. The phantom analysis found no significant difference between the left- and right-side scans over time for length (p=0.224) and for thickness (p=0.314), and an accuracy of 99.58% and 89.04%, respectively. When comparing the phantom scans over time, a consistent Hausdorff distance of 3.2 mm was found between phantom scans. A significant difference was found for the estimated width when comparing the patient DBT to SL data (p=0.004), but this was not the case for CND (p=0.110). To conclude, the SL scanning system has a high precision and is accurate to within approximately 3 mm compared to the ground truth.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
F. Schampers, M. C. Pinto, and K. Michielsen "Accuracy and precision of a structured light scanning system for acquiring 3D information of the breast under compression", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 115954T (15 February 2021); https://doi.org/10.1117/12.2581913
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

3D image processing

Structured light

Digital breast tomosynthesis

Digital mammography

Image compression

Image processing