We investigated the spectral performance of pnCCDs using a small-sized device mounted on the camera system manufactured by PNDetector GmbH. Several fluorescence X-ray lines from 0.28 keV(C-Kα) to 4.5keV (Ti-Kα) were used for measurements. The energy resolution of 52 eV for 0.28 keV was verified, with a frame rate of 100 Hz at −20 °C of operating temperature. We also made measurements of dark current for various temperatures ranging from −20 to 0 °C and evaluated that the lower discriminated energy could be 123 eV even at 0 °C. These results verified to satisfy the mission requirements of HiZ-GUNDAM. |