Paper
1 November 1990 Two-dimensional encoding of Raman emission for detection with a charge-coupled device
Michael J. Pelletier
Author Affiliations +
Abstract
This report investigates the performance of a CCD-echelle spectrograph designed for Raman spectroscopy. A holographic band-reject filter is used to reject elastically scattered light without seriously reducing the instrument throughput. The capabilities of this instrument are demonstrated by a cyclohexane spectrum collected in 300 milliseconds with 12 milliwatts of laser power, and by a spectrum of 1% L-phenylalanine in water. Index Headings: Raman spectroscopy, Charge-coupled device, Echelle, Holographic filter
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michael J. Pelletier "Two-dimensional encoding of Raman emission for detection with a charge-coupled device", Proc. SPIE 1336, Raman and Luminescence Spectroscopies in Technology II, (1 November 1990); https://doi.org/10.1117/12.22905
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Raman spectroscopy

Spectrographs

Charge-coupled devices

Optical filters

Holography

CCD cameras

CCD image sensors