1 June 1991 High-resolution image digitizing camera for use in quantitative coronary arteriography
Markus H. Muser, Thomas Leemann, M. Anliker
Author Affiliations +
Proceedings Volume 1448, Camera and Input Scanner Systems; (1991) https://doi.org/10.1117/12.45349
Event: Electronic Imaging '91, 1991, San Jose, CA, United States
Abstract
Image processing in biomedical applications such as the analysis of electrophoresis gels, digital microscopic imaging or the computer-assisted quantitative analysis of angiographic images recorded on 35 mm cinefilm quantitative coronary arteriography (QCA) can be improved substantially if the conventional TV-based image digitizers are replaced by devices offering a higher geometric resolution and an increased dynamic range. Before high resolution two-dimensional CCD Sensors were introduced a few years ago, these improvements could only be implemented cost-effectively by using scanning devices such as the digital camera developed earlier at the Institute for Biomedical Engineering in Zurich, Switzerland. This camera, based on a 2048-element CCD line array, offers a geometrical resolution of up to 2048 X 3000 pixel and a dynamic range of up to 12 bit. It is being used in the QCA system developed jointly by the Institute for Biomedical Engineering in Zurich, Switzerland and the University of Texas Health Science Center in Houston, Texas, U.S.A. Clinical evaluation of this system as well as an analysis of the technical properties of the X-Ray systems involved indicate that coronary arteriograms can be digitized with a resolution of about 1024 X 1024 pixel without sacrificing measurement accuracy. This fact reduces computational effort and suggests the use of two-dimensional CCD sensors. Therefore, a new digitizing camera based on a KAF1400 (Kodak) full frame sensor suitable for the QCA system has been developed. Its design concept and performance are discussed in the paper.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Markus H. Muser, Thomas Leemann, and M. Anliker "High-resolution image digitizing camera for use in quantitative coronary arteriography", Proc. SPIE 1448, Camera and Input Scanner Systems, (1 June 1991); https://doi.org/10.1117/12.45349
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Sensors

Imaging systems

Angiography

Image resolution

X-rays

Digital cameras

RELATED CONTENT

Development Of A High Resolution X Ray Imaging Device For...
Proceedings of SPIE (January 01 1987)
The X-Ray Imaging System
Proceedings of SPIE (June 01 1972)
Z/I Imaging digital aerial camera system
Proceedings of SPIE (November 29 2000)
X Ray Image Sensor Based On An Optical TDI CCD...
Proceedings of SPIE (May 01 1986)

Back to Top