1 October 1991 UCSD high-energy x-ray timing experiment magnetic shield design and test results
Author Affiliations +
Abstract
Results are reported from an effort to define a passive magnetic field concept for the High Energy X-ray Timing Experiment (HEXTE), in the interest of reducing the detector-gain variations due to 0.5-1.0-sec timescale magnetic field variations. This will allow a sensitivity of the order of 1 percent of the HEXTE background. While aperture modulation and automatic gain control will minimize effects on timescales of tens of seconds and longer, passive magnetic shielding of the photomultiplier tubes will address 1-sec timescale variations due to aperture motions.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Richard E. Rothschild, Michael R. Pelling, and Paul L. Hink "UCSD high-energy x-ray timing experiment magnetic shield design and test results", Proc. SPIE 1549, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy II, (1 October 1991); https://doi.org/10.1117/12.48333