You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
14 September 1993Performance characteristics of a flexible antiscatter grid used in computed radiography
In this paper we examine the imaging performance of an experimental mechanically-flexible antiscatter grid for use with computed radiography. Bucky factors, contrast improvement factors, and signal-to-noise improvement factors were computed from transmittance measurements made with 5-, 10-, 15-, 20-, and 25-cm uniform-thickness water phantoms. These phantoms were selected to span the range of scatter fractions found in chest radiography.
The alert did not successfully save. Please try again later.
Richard L. Van Metter, David P. Trauernicht, David J. Steklenski, "Performance characteristics of a flexible antiscatter grid used in computed radiography," Proc. SPIE 1896, Medical Imaging 1993: Physics of Medical Imaging, (14 September 1993); https://doi.org/10.1117/12.154600