You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
1 May 1994Direct x-ray sensing CCD array for intraoral dental x-ray imaging system
A commercial prototype electronic intraoral dental x-ray imaging system employing a direct sensing CCD array has been developed. Image quality parameters were measured using x-ray sources at the National Institute of Standard and Technology radiation physical department in Gaithersburg, MD. Detector response to x-rays in the 10 to 70 keV energy range was measured. The beam hardening effects of human anatomy on a typical 70 kVp spectra was measured using a tissue-equivalent dental phantom.
The alert did not successfully save. Please try again later.
John D. Cox, Donald W. Williams, D. Scott Langford, "Direct x-ray sensing CCD array for intraoral dental x-ray imaging system," Proc. SPIE 2163, Medical Imaging 1994: Physics of Medical Imaging, (1 May 1994); https://doi.org/10.1117/12.174266