This paper represents the method of contact zone effective temperature measuring by registering thermal radiation which is back propagated through the main laser radiation delivery fiber channel. The results of contact zone temperature measuring experiments for some different tissue types are also represented. The present method allows us to obtain a reliable real-time information about instant temperature in the contact zone using a reliable scale of time. The possibility of thermal feed-back control of an in contact laser destructor based on Ho laser is evaluated.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gregory B. Altshuler, Andrew V. Erofeev, Ildar K. Ilyasov, and Constantin V. Prikhodko "Local fiber tip temperature monitoring for the Ho:YAG contact laser scalpel", Proc. SPIE 2327, Medical Applications of Lasers II, (23 December 1994); https://doi.org/10.1117/12.197560