At present a number of ozone analyzer types with their own advantages and disadvantages are produced in Russia. Being wide in use the chemiluminescent ozone analyzers need a calibrator (ozone generator) with specified ozone concentration to compare with ozone concentration measured in the atmosphere. The authors describe a 2-beams UV photometer to graduate ozone generators. The instrument allows to measure background ozone concentrations in the ground layer of atmosphere. It has two equal absorption chambers (cells) switched alternately by two gas solenoid valves from the reference to measurement modes. The concentration value is determined by discrepancy.
Vladimir A. Polischyuk,Sergey I. Krasko,Mikhail B. Levin, andVladimir P. Chelibanov
"Simple photometer of F103 model: a stand-alone unit for ozone checking gas mixture generators", Proc. SPIE 2503, Air Toxics and Water Monitoring, (21 September 1995); https://doi.org/10.1117/12.221095