Paper
3 January 1996 Fiber optic medical pressure-sensing system employing intelligent self-calibration
Author Affiliations +
Abstract
In this article, we describe a fiber-optic catheter-type pressure-sensing system that has been successfully introduced for medical diagnostic applications. We present overall sensors and optoelectronics designs, and highlight product development efforts that lead to a reliable and accurate disposable pressure-sensing system. In particular, the incorporation of an intelligent on-site self-calibration approach allows limited sensor reuses for reducing end-user costs and for system adaptation to wide sensor variabilities associated with low-cost manufacturing processes. We demonstrate that fiber-optic sensors can be cost-effectively produced to satisfy needs of certain medical market segments.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gang He "Fiber optic medical pressure-sensing system employing intelligent self-calibration", Proc. SPIE 2594, Self-Calibrated Intelligent Optical Sensors and Systems, (3 January 1996); https://doi.org/10.1117/12.229232