Paper
15 January 1996 Laser interferometric vibration measurements of the middle ear in healthy humans
J. Rodriguez Jorge, Werner Hemmert, Claus Burkhardt, Hans-Peter Zenner, Anthony W. Gummer
Author Affiliations +
Abstract
The use of spontaneous and evoked otacoustic emissions is now a standard clinical tool for diagnosis of the function of the inner ear. However, it is not possible to extract this information over the entire, functionally relevant frequency range because of imperfect coupling of: (1) stapedial to ear-drum vibrations through the ossicular chain of the middle ear and (2) ear-drum vibrations to air in the external auditory meatus. The problem could be circumvented if it were possible to measure the vibration of the stapes and ear drum. The ear drum can be visualized non-invasively, whereas the stapes is only accessible intra-operatively. Therefore, we designed a laser-interferometric system to non-invasively measure the vibration of the human ear drum. Vibrations were measured with a laser Doppler velocimeter (Polytec OFV-302) coupled into the side arm of an operating microscope (Zeiss OPMI MDM). The wavelength was 633 nm and emitted power was less than 1 mW. Direct coupling through the optics of the operating microscope, instead of through glass fibers, enabled a larger signal-to- noise ratio (20 - 30 dB) due to collection of more reflected light. This coupling scheme avoids the problems associated with having to place a reflecting material on the ear drum. The developed vibration measurement system allows non-invasive, fast and reproducible characterization of the dynamics of the human ear drum and as such can be used for clinical diagnostics.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Rodriguez Jorge, Werner Hemmert, Claus Burkhardt, Hans-Peter Zenner, and Anthony W. Gummer "Laser interferometric vibration measurements of the middle ear in healthy humans", Proc. SPIE 2628, Optical and Imaging Techniques for Biomonitoring, (15 January 1996); https://doi.org/10.1117/12.229985
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ear

Acoustics

Scanning probe lithography

Microscopes

Doppler effect

Vibrometry

Signal to noise ratio