21 May 1996 Calibration of scanning electron microscope magnification standards SRM484
Joseph Fu, Theodore V. Vorburger, David B. Ballard
Author Affiliations +
Abstract
Standard Reference Material (SRM) 484 is an artifact for calibrating the magnification scale of a scanning electron microscope. Since 1977 the National Institute ofStandards and Technology (MST) has produced seven issues of SRM484 amounting to approximately 1 150 samples in all. The standards are fabricated by electroplating alternate layers ofnickel and gold onto a substrate of a Monel sheet metal. The plate is then diced, and the individual pieces are mounted on edge in a holder. Each sample is metallographically polished to obtain a smooth surface and to reveal gold lines. The samples are calibrated using a scanniig electron microscope incorporated with a laser interferometer. A piezo flexure stage carries the sample across the stationary electron beam. A backscattered electron detector detects an intensity peak at each location where the electron beam interacts with a gold line. The displacement ofthe traveled stage between lines is monitored by the interferometer. A computer program records the intensity peaks and displacement information and determines the distance (spacing) between any two peaks. The spacing is measured from a peak-to-peak algorithm rather than an edge-to-edge algorithm in order to avoid the determination ofline edge positions. Properties ofthe SRM484 and the measurement system result in recent expanded uncertainties (at the level oftwo standard deviations) of approximately 4% for 0.5 jim spacings and 0.5% for 50 tm spacings. Key words: Interferometer, Measurement uncertainty, Scanning Electron Microscope, SEM magnification, Standard Reference Material.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joseph Fu, Theodore V. Vorburger, and David B. Ballard "Calibration of scanning electron microscope magnification standards SRM484", Proc. SPIE 2725, Metrology, Inspection, and Process Control for Microlithography X, (21 May 1996); https://doi.org/10.1117/12.240150
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Scanning electron microscopy

Electron microscopes

Interferometers

Gold

Electron beams

Microscopes

Back to Top