Abstract
The use of information contained in the transmitted x-ray spectrum provides the capability for selective removal of substances of a particular mean atomic number in projection radiographs. Using a prototype system for line-scanned digital radiography, x-ray images were produced with two different x-ray spectra by modulating the kVp applied to the x-ray source and filtering the x-ray beam. Using a previously described Compton/photoelectric decomposition algorithm, subtraction images are made with bone or soft-tissue (water) shadows removed. Applications to chest, abdominal and skeletal imaging in-vivo are demonstrated.
© (1981) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
William R. Brody, F. Graham Sommer, Leonard A Lehmann, Albert Macovski, Robert E. Alvarez, Norbert J. Pelc, Stephen J. Riederer, and Anne Hall "Dual-kVp Radiography", Proc. SPIE 0273, Application of Optical Instrumentation in Medicine IX, (16 July 1981); https://doi.org/10.1117/12.931810
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiography

Bone

Dual energy imaging

Tissues

X-rays

X-ray imaging

Visualization

Back to Top