8 November 1996 High-accuracy characterization and applications of filter radiometers
Nigel P. Fox, Christopher J. Chunnilall, Neil J. Harrison, William S. Hartree
Author Affiliations +
Abstract
The paper describes the high-accuracy radiometric calibration of filter radiometers using both laser and Fourier transform spectrometer based methods to uncertainty levels < 0.1%. The paper also describes the development of detector-based transfer standards and their use for the spectral calibration of radiometric sources. It discusses how these techniques can improve the accuracy and reduce the cost of source calibration.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Nigel P. Fox, Christopher J. Chunnilall, Neil J. Harrison, and William S. Hartree "High-accuracy characterization and applications of filter radiometers", Proc. SPIE 2815, Optical Radiation Measurements III, (8 November 1996); https://doi.org/10.1117/12.257158