Paper
28 May 1997 Improvements of UV/x-ray framing image tube cameras
J. Feng, Yaonan K. Ding, Zuli Liu, Qin Lao Yang, B. P. Guo, Hanben Niu
Author Affiliations +
Proceedings Volume 2869, 22nd International Congress on High-Speed Photography and Photonics; (1997) https://doi.org/10.1117/12.273466
Event: 22nd International Congress on High-Speed Photography and Photonics, 1996, Santa Fe, NM, United States
Abstract
The framing image tube camera has some intrinsic advantages. It has got some applications in ICF and X-ray laser field. But there are two main problems with this technique: the loss of spatial resolution at short exposure time,a nd the electromagnetic disturbance in the experimental area. In order to solve these problems, we have taken several measures: to improve the design of the image tube and the electronics and to strengthen the shielding. Additionally, in intense X-ray pulse we use MgF2 as cathode material instead of Au and CsI to reduce the space charge effects. The experiments in high power laser area have shown that 3 frame images can be provided with 75 ps (FWHM) exposure time and dynamic spatial resolution of 6 lp/mm. Some physics results of high power laser-plasma interaction have also been obtained.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Feng, Yaonan K. Ding, Zuli Liu, Qin Lao Yang, B. P. Guo, and Hanben Niu "Improvements of UV/x-ray framing image tube cameras", Proc. SPIE 2869, 22nd International Congress on High-Speed Photography and Photonics, (28 May 1997); https://doi.org/10.1117/12.273466
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Spatial resolution

X-rays

High power lasers

Magnesium fluoride

Electromagnetism

Gold

RELATED CONTENT

Picosecond Soft X-Ray Streak Camera
Proceedings of SPIE (June 07 1989)
High-speed x-ray imaging in high-power laser experiments
Proceedings of SPIE (March 17 2005)
Picosecond X-Ray Chronoscopy
Proceedings of SPIE (September 14 1977)
Comparison of the uniformity of soft and hard x ray...
Proceedings of SPIE (October 14 1997)
High Speed (<250 ps) High Gain X-Ray Shutter Camera
Proceedings of SPIE (March 01 1983)

Back to Top