You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
2 May 1997Quasi-monochromatic x-ray source for mammography via crystal array
It has been demonstrated both theoretically and experimentally that mammographic x-ray imaging with monochromatic beams could help to improve subject contrast and to reduce the dose delivered to the patient. To this aim, quasi-monochromatic x-rays have been produced in the mammographic energy range by making use of a conventional W- anode, Be-window x-ray tube and a monochromator optical system based on an array of mosaic crystals. The mosaic crystals are highly oriented pyrolytic graphite which provide a gain in flux as compared to perfect crystals because of their higher integrated reflectivity. The monochromator optical system consists of an array of three crystals which has been assembled so as to produce in the image plane an irradiation field obtained with adjacent reflected beams. The field size reflected by each crystal on the image plane is limited by the desired energy resolution along the horizontal direction and by the crystal size along the perpendicular one. The energy spread of the reflected beams is about 10 percent. The characteristics of the system in terms of energy resolution and fluence rate are reported. Radiographs of test phantoms imaged with quasi-monochromatic beams in the energy range of 18-21 keV have been obtained with a conventional screen/film combination. To remove the spatial non-uniformities of the entire irradiation field a correction procedure has been applied.Large field quasi- monochromatic x-ray beams with the same flux of a standard Mo-anode tube with an anode current of about 600 mA.
The alert did not successfully save. Please try again later.
Mauro Gambaccini, Alberto Del Guerra, Angelo Taibi, Alessandra Tuffanelli, "Quasi-monochromatic x-ray source for mammography via crystal array," Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.273980