15 October 1997 Hard x-ray calibration facility design for JEM-X detector on board INTEGRAL
Giovanni Pareschi, Filippo Frontera, Carlo Pelliciari, Guido Zavattini, Vittore Carassiti, Federico Evangelisti
Author Affiliations +
Abstract
The x-ray monitor JEM-X will be flown aboard the ESA mission INTEGRAL for gamma-ray astronomy. It has been conceived to furnish images of celestial sources in fields of the sky 9.1 deg wide in the 3 - 100 keV extended energy band. Careful on- ground calibrations are needed to verify the spectral capabilities of the JEM-X detector. These operations in the 20 - 100 keV energy band are planned to be performed using a hard x-ray (15 - 140 keV) facility operated at University of Ferrara (Italy). We present the apparatus that will be used to carry out these calibrations. In particular we describe a double-reflection fixed-exit system of monochromatization based on silicon crystals that are implemented in the facility.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Giovanni Pareschi, Filippo Frontera, Carlo Pelliciari, Guido Zavattini, Vittore Carassiti, and Federico Evangelisti "Hard x-ray calibration facility design for JEM-X detector on board INTEGRAL", Proc. SPIE 3114, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy VIII, (15 October 1997); https://doi.org/10.1117/12.283783
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Crystals

Calibration

X-rays

Sensors

Monochromators

Laser crystals

Hard x-rays

RELATED CONTENT

Performance of the x ray facility for calibration of the...
Proceedings of SPIE (October 22 1999)
High-accuracy X-ray detector calibration at PTB
Proceedings of SPIE (September 29 2004)
Study and implementation of a soft x ray 100 eV...
Proceedings of SPIE (June 27 2006)

Back to Top