Paper
4 November 1981 Digital Imaging For Radiation Therapy Verification
S. C. Taborsky, W. C. Lam, R. E. Sterner, G. M. Skarda
Author Affiliations +
Abstract
A linear diode array was used to scan a radiation field for pulsed high energy x-ray imaging. With a signal of about 50 pico-coulombs of charge from each diode for each beam pulse, the signal to noise ratio for an image of an open field was 119. An image of a porcine femur in a wax phantom was obtained by scanning the transmitted radiation. The resultant electronic image was superior in quality when compared to the standard radiotherapy port film. The simplicity of the detector system has demonstrated the feasibility of constructing an electronic portal imaging system for clinical use.
© (1981) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. C. Taborsky, W. C. Lam, R. E. Sterner, and G. M. Skarda "Digital Imaging For Radiation Therapy Verification", Proc. SPIE 0314, Digital Radiography, (4 November 1981); https://doi.org/10.1117/12.933040
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Diodes

Detector arrays

Signal detection

Signal to noise ratio

Photons

Bone

Back to Top