Paper
5 June 1998 Glucose measurements in solutions using fiber optic evanescent wave spectroscopy and tunable CO2 laser
Yaron Gotshal, Ido Adam, Abraham Katzir
Author Affiliations +
Proceedings Volume 3262, Surgical-Assist Systems; (1998) https://doi.org/10.1117/12.309483
Event: BiOS '98 International Biomedical Optics Symposium, 1998, San Jose, CA, United States
Abstract
There is an interest in using IR spectroscopy techniques for blood analysis, and in particular for glucose measurements. Glucose concentration in aqueous solutions were measured using fiberoptic evanescent wave spectroscopy. A new spectrometer consisting of tunable CO2 laser as the radiation source, and a silver halide fiber was built. Several solutions were measured for calibration with concentrations in the range of 70-2000 mg/dl. After the calibration, solution were measured and predicted according to the calibration parameters. The predicted concentrations were in agreement with the reference ones. The predicted errors for concentrations range up to 200 mg/dl, which is the clinical range, was less then 18 mg/dl (a relative error of about 12%). The AgClBr fibers are non-toxic and therefore this measurement can be carried out in real-time, for in-situ measurements of glucose in blood.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yaron Gotshal, Ido Adam, and Abraham Katzir "Glucose measurements in solutions using fiber optic evanescent wave spectroscopy and tunable CO2 laser", Proc. SPIE 3262, Surgical-Assist Systems, (5 June 1998); https://doi.org/10.1117/12.309483
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Glucose

Spectroscopy

Carbon dioxide lasers

Sensors

Blood

Calibration

Fiber optics

Back to Top