26 October 1998 Results of a pyroelectric detector calibration by a cryogenic radiometer at 10.6 mm
Jeanne M. Houston, Thomas R. Gentile
Author Affiliations +
Abstract
A pyroelectric detector was calibrated against the National Institute of Standards and Technology (NIST) High Accuracy Cryogenic Radiometer (HACR) at 10.6 mm using a CO2 laser as a source. The purpose of this calibration was to link the NIST infrared (IR) radiometric detector scales to the HACR. Issues addressed in this calibration included the spatial non- uniformity of the pyroelectric detector, the transport and alignment of the IR laser beam, and the measurement of ac signals. The final combined relative standard uncertainty of the calibration is 0.48%, with the largest uncertainty component arising from the spatial non-uniformity of the pyroelectric detector. The apparatus, measurement procedures, and results of the calibration will be discussed.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jeanne M. Houston and Thomas R. Gentile "Results of a pyroelectric detector calibration by a cryogenic radiometer at 10.6 mm", Proc. SPIE 3436, Infrared Technology and Applications XXIV, (26 October 1998); https://doi.org/10.1117/12.327991
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Sensors

Pyroelectric detectors

Cryogenics

Radiometry

Radio optics

Standards development

RELATED CONTENT

How to obtain traceability on optical radiation measurements?
Proceedings of SPIE (February 10 2006)
Low NEP pyroelectric radiometer standards
Proceedings of SPIE (May 01 2008)
Limits of radiometry
Proceedings of SPIE (October 17 1994)

Back to Top