Ping Zhao, Robert A. Austin, Richard J. Edgar, Ronald F. Elsner, Terrance J. Gaetz, Dale E. Graessle, Diab Jerius, Jeffery J. Kolodziejczak, Walter C. McDermott, Stephen L. O'Dell, Martin E. Sulkanen, Daniel A. Schwartz, Douglas A. Swartz, Allyn F. Tennant, Leon P. Van Speybroeck, Bradford J. Wargelin, Martin C. Weisskopf, C. G Zirnstein
Author Affiliations +
Abstract
The AXAF X-ray mirrors underwent thorough calibration using the X-ray Calibration Facility (XRCF) at the Marshall Space Flight Center in Huntsville, AL from late 1996 to early 1997. The x-ray calibration made novel use of the x-ray continuum from a conventional electron-impact source. Taking advantage of the good spectral resolution of solid-state detectors, continuum measurements proved advantageous in calibration the effective area of AXAF's High-Resolution Mirror Assembly (HRMA) for the entire AXAF energy band. The measurements were made by comparing the spectrum detected by a SSD at the focal plane with the spectrum detected by a beam normalization SSD. The HRMA effective area was calibrated by comparing the measurements with the HRMA raytrace model. The HRMA on-orbit performance predictions are made using the calibration results.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ping Zhao, Robert A. Austin, Richard J. Edgar, Ronald F. Elsner, Terrance J. Gaetz, Dale E. Graessle, Diab Jerius, Jeffery J. Kolodziejczak, Walter C. McDermott, Stephen L. O'Dell, Martin E. Sulkanen, Daniel A. Schwartz, Douglas A. Swartz, Allyn F. Tennant, Leon P. Van Speybroeck, Bradford J. Wargelin, Martin C. Weisskopf, and C. G Zirnstein "AXAF-mirror effective area calibration using the C-continuum source and solid state detectors", Proc. SPIE 3444, X-Ray Optics, Instruments, and Missions, (19 November 1998); https://doi.org/10.1117/12.331240
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

X-rays

Sensors

Quantum efficiency

Mirrors

Error analysis

Data modeling

RELATED CONTENT

HETG high-order diffraction efficiency
Proceedings of SPIE (November 19 1998)
Effective area of the AXAF high-resolution camera (HRC)
Proceedings of SPIE (November 19 1998)
X-ray CCD calibration for the AXAF CCD Imaging Spectrometer
Proceedings of SPIE (October 31 1996)
Calibrating the wings of the Chandra PSF
Proceedings of SPIE (February 03 2004)
Correcting x ray spectra obtained from the AXAF VETA I...
Proceedings of SPIE (January 21 1993)

Back to Top