Translator Disclaimer
Paper
9 July 1999 Enhancing the contrast of subcutaneous veins
Author Affiliations +
Proceedings Volume 3595, Biomedical Diagnostic, Guidance, and Surgical-Assist Systems; (1999) https://doi.org/10.1117/12.351535
Event: BiOS '99 International Biomedical Optics Symposium, 1999, San Jose, CA, United States
Abstract
A technique for enhancing the contrast of subcutaneous veins has been demonstrated. This technique uses a near infrared light source and one or more infrared sensitive CCD TV cameras to produce a contrast enhanced image of the subcutaneous veins. This video image of the veins is projected back onto the patient's skin using an LCD vein projector. The use of an infrared transmitting filter in front of the video cameras prevents any positive feedback from the visible light from the video projector from causing instabilities in the projected image. The demonstration contrast enhancing illuminator has been tested on adults, both Caucasian and African-American, and it enhances veins quite well in most cases. Preliminary studies on a 9 month old girl indicate promise for pediatric use.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Herbert D. Zeman and Gunnar Lovhoiden "Enhancing the contrast of subcutaneous veins", Proc. SPIE 3595, Biomedical Diagnostic, Guidance, and Surgical-Assist Systems, (9 July 1999); https://doi.org/10.1117/12.351535