Paper
20 May 1999 CT artifacts in virtual colonoscopy
Bruce R. Whiting, Elizabeth G. McFarland, Casey Dellabarca, Barry S. Brunsden
Author Affiliations +
Abstract
A phantom representation of typical colon structures with precisely known geometrical measurements was designed and fabricated. Computed tomography (CT) data were collected using a range of protocols typical for spiral CT colonography. Analysis methods were developed to measure the acquired geometry of the phantom data and characterize distortions/degradation. Simple models were proposed to explain the trends in degradation in the acquisition process versus scanner protocol. Preliminary results indicate that degradation due to CT acquisition will not significantly impact the detection of clinically relevant lesions (dimensions greater than 1 cm). However, the CT acquisition process does place a lower limit on detection size of several millimeters.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bruce R. Whiting, Elizabeth G. McFarland, Casey Dellabarca, and Barry S. Brunsden "CT artifacts in virtual colonoscopy", Proc. SPIE 3660, Medical Imaging 1999: Physiology and Function from Multidimensional Images, (20 May 1999); https://doi.org/10.1117/12.349579