22 September 1999 Developments in synchrotron x-ray computed microtomography at the National Synchrotron Light Source
Betsy A. Dowd, Graham H. Campbell, Robert B. Marr, Vivek V. Nagarkar, Sameer V. Tipnis, Lisa Axe, D. Peter Siddons
Author Affiliations +
Abstract
Last year, the X27A beamline at the National Synchrotron Light Source (NSLS) became dedicated solely to X-Ray Computed Microtomography (XCMT). This is a third-generation instrument capable of producing tomographic volumes of 1 - 2 micron resolution over a 2 - 3 mm field of view. Recent enhancements will be discussed. These have focused on two issues: the desire for real-time data acquisition and processing and the need for highly monochromatic beam (.1% energy bandpass). The latter will permit k-edge subtraction studies and will provide improved image contrast from below the Cr (6 keV) up to the Cs (36 keV) k-edge. A range of applications that benefit from these improvements will be discussed as well. These two goals are somewhat counterproductive, however; higher monochromaticity yields a lower flux forcing longer data acquisition times. To balance the two, a more efficient scintillator for X-ray conversion is being developed. Some testing of a prototype scintillator has been performed; preliminary results will be presented here. In the meantime, data reconstruction times have been reduced, and the entire tomographic acquisition, reconstruction and volume rendering process streamlined to make efficient use of synchrotron beam time. A Fast Filtered Back Transform (FFBT) reconstruction program recently developed helped to reduce the time to reconstruct a volume of 150 X 150 X 250 pixels3 (over 5 million voxels) from the raw camera data to 1.5 minutes on a dual R10,000 CPU. With these improvements, one can now obtain a 'quick look' of a small tomographic volume (approximately 106 voxels) in just over 15 minutes from the start of data acquisition.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Betsy A. Dowd, Graham H. Campbell, Robert B. Marr, Vivek V. Nagarkar, Sameer V. Tipnis, Lisa Axe, and D. Peter Siddons "Developments in synchrotron x-ray computed microtomography at the National Synchrotron Light Source", Proc. SPIE 3772, Developments in X-Ray Tomography II, (22 September 1999); https://doi.org/10.1117/12.363725
Lens.org Logo
CITATIONS
Cited by 171 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scintillators

X-rays

Data acquisition

Charge-coupled devices

Synchrotrons

Monochromators

Tomography

RELATED CONTENT

High-resolution large-area high-energy x-ray tomography
Proceedings of SPIE (October 24 1997)
Sensitivity of x ray phase tomography based on Talbot and...
Proceedings of SPIE (September 17 2008)
X-ray tomographic microscopy at the Swiss Light Source
Proceedings of SPIE (January 07 2002)
Operational experience of a large area x ray camera for...
Proceedings of SPIE (October 06 1999)
X-ray detector for time-resolved studies
Proceedings of SPIE (November 01 1991)

Back to Top