6 October 1999 Performance of multiple-CCD x-ray detector system at SPring-8 for protein crystallography
Masayo Suzuki, Masaki Yamamoto, Takashi Kumasaka, Kazumichi Sato, Hidenori Toyokawa, Tatzuo Ueki
Author Affiliations +
Abstract
The 4 X 4 multiple CCD X-ray detector installed in the beamline 45XU of the SPring-8 facility in Japan was examined by acquiring X-ray diffraction images of lysozyme crystals. The detector system was calibrated at the beamline by observing X-ray images of a flood-field and a mask pattern generated by the synchrotron radiation X-ray beam. It was found in the present work that the quality of the X-ray images observed is limited by an unexpected spike noise, the origin of which was suspected to be the (gamma) -rays emitted from the radioisotopes remaining in the X-ray scintillating screen. Image analysis carried out with a median filter predicted, however, that the image quality could be as high as designed, once the noise as well as the grain colonization is removed from the screen.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Masayo Suzuki, Masaki Yamamoto, Takashi Kumasaka, Kazumichi Sato, Hidenori Toyokawa, and Tatzuo Ueki "Performance of multiple-CCD x-ray detector system at SPring-8 for protein crystallography", Proc. SPIE 3774, Detectors for Crystallography and Diffraction Studies at Synchrotron Sources, (6 October 1999); https://doi.org/10.1117/12.367121
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

X-ray detectors

Charge-coupled devices

X-ray imaging

Sensors

Digital filtering

Image filtering

Back to Top