Paper
17 April 2001 Possible application of plasma x-ray source to clinical radiography
Yukiko Nakajima, Eiichi Sato, Kenkichi Tanioka, Toshiyuki Kawai, Etsuro Tanaka, Waka Kobayashi, Yutaka Yamamoto, Mototsugu Fujii, Yoshiro Shinozaki, Yoshio Kan, Kazuyuki Hyodo, Masami Ando, Hidezo Mori
Author Affiliations +
Proceedings Volume 4183, 24th International Congress on High-Speed Photography and Photonics; (2001) https://doi.org/10.1117/12.424300
Event: 24th International Congress on High-Speed Photography and Photonics, 2000, Sendai, Japan
Abstract
Small amount of iodine contrast material in the small vessels can be efficiently detected by monochromatic synchrotron radiation (x-ray) tuned at just above k-edge of iodine (33 KeV). However, the serious disadvantage of the system is a high cost for the facility to obtain synchrotron radiation (x- ray). We are trying to apply a plasma x-ray source for this purpose. In this system, the high voltage main condenser is charged up to 60 KeV by a power supply. The electric charges in the condenser are discharged to the tube after triggering the cathode electrode. For a cerium plasma target, a highly intensified K(alpha) and K(beta) characteristic x-ray of cerium (34.6 KeV and 38.3 KeV) was produced, as bremsstrahlung x-ray was absorbed and converted into fluorescent x-ray. Plasma x- ray source might be an alternative of monochromatic synchrotron radiation for small vessel radiography.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yukiko Nakajima, Eiichi Sato, Kenkichi Tanioka, Toshiyuki Kawai, Etsuro Tanaka, Waka Kobayashi, Yutaka Yamamoto, Mototsugu Fujii, Yoshiro Shinozaki, Yoshio Kan, Kazuyuki Hyodo, Masami Ando, and Hidezo Mori "Possible application of plasma x-ray source to clinical radiography", Proc. SPIE 4183, 24th International Congress on High-Speed Photography and Photonics, (17 April 2001); https://doi.org/10.1117/12.424300
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Synchrotron radiation

Plasma

Arteries

X-ray sources

Tumors

Heart