You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
28 May 2001Evaluation of a new method for stenosis quantification from 3D x-ray angiography images
A new method for stenosis quantification from 3D X-ray angiography images has been evaluated on both phantom and clinical data. On phantoms, for the parts larger or equal to 3 mm, the standard deviation of the measurement error has always found to be less or equal to 0.4 mm, and the maximum measurement error less than 0.17 mm. No clear relationship has been observed between the performances of the quantification method and the acquisition FoV. On clinical data, the 3D quantification method proved to be more robust to vessel bifurcations than its 3D equivalent. On a total of 15 clinical cases, the differences between 2D and 3D quantification were always less than 0.7 mm. The conclusion is that stenosis quantification from 3D X-4ay angiography images is an attractive alternative to quantification from 2D X-ray images.
The alert did not successfully save. Please try again later.
Fabienne Betting, Gilles Moris, Jerome Knoplioch, Yves L. Trousset, Francisco Sureda, Laurent Launay, "Evaluation of a new method for stenosis quantification from 3D x-ray angiography images," Proc. SPIE 4319, Medical Imaging 2001: Visualization, Display, and Image-Guided Procedures, (28 May 2001); https://doi.org/10.1117/12.428054