Paper
30 May 2001 Ultrasonic strain imaging in media with pulsatile flow
Jerome J. Mai, Christian M. Kargel, Hisham Mhanna, Michael F. Insana
Author Affiliations +
Abstract
This report summarizes our experiments with novel ultrasound phantoms that mimic essential bio mechanical and dynamical vascular features of soft biological tissues. Real-time RF echo acquisition using a Siemens Elegra ultrasound system at 7.5 MHz provided a time series of ultrasonic images that were used to image longitudinal strain from pulsed and steady flows. Physical features of internal deformation patterns resulting from pulsatile flows revealed that ultrasonic strain imaging could be a very sensitive method for observing important properties related to physiological fluid flow.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jerome J. Mai, Christian M. Kargel, Hisham Mhanna, and Michael F. Insana "Ultrasonic strain imaging in media with pulsatile flow", Proc. SPIE 4325, Medical Imaging 2001: Ultrasonic Imaging and Signal Processing, (30 May 2001); https://doi.org/10.1117/12.428190
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissues

Ultrasonography

Ultrasonics

Data acquisition

Image compression

Image processing

Acoustics