3 May 2002 Anti-aliasing weighting functions for multislice helical CT
Author Affiliations +
Abstract
We develop a new projection weighting function for interpolation and reconstruction of multi-slice helical computed tomography data with the hope of reducing longitudinal aliasing in reconstructed volumes. The weighting function is based on the application of the Papoulis generalized sampling theorem to the interlaced longitudinal samples acquired by the multi-slice scanner. We call the approach 180MAA, for multi-slice anti-aliasing. For pitch 3, the 180MAA approach yields high-quality images of the 3D Shepp-Logan phantom as well as a longitudinal MTF superior to that of the 180MLI approach, which is based on the use of linear interpolation. However, it is not as successful at mitigating aliasing as had been doped due to the presence of a significant and unexpected aliasing component that can be attributed to the small cone angle in multi-slice helical CT. The presence of this effect is interesting and significant in its own right, however.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Patrick J. La Riviere and Xiaochuan Pan "Anti-aliasing weighting functions for multislice helical CT", Proc. SPIE 4682, Medical Imaging 2002: Physics of Medical Imaging, (3 May 2002); https://doi.org/10.1117/12.465580