We report on ultrahigh-resolution Optical Coherence Tomography (OCT) using white-light interference microscopy. The experimental setup is based on a Linnik interferometer illuminated by a tungsten halogen lamp. Tomographic images in the en face orientation are calculated by combination of images recorded by a silicon CCD camera. Axial resolution of 0.8 μm is achieved due to the short coherence length of the source and the compensation of dispersion mismatch in the interferometer arms. Transverse resolution of 1.6 μm is obtained by using relatively high numerical aperture microscope objectives. A nearly shot-noise limited detection sensitivity of 90 dB is achieved with 4s acquisition time. Images of the Xenopus Laevis tadpole are presented.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Arnaud Dubois, Laurent Vabre, Romain Lecaque, and Albert-Claude Boccara "Ultrahigh-resolution OCT using white-light interference microscopy", Proc. SPIE 4956, Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine VII, (8 July 2003); https://doi.org/10.1117/12.477631
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 3 patents.